Dose profile
In External Beam Radiotherapy, transverse and longitudinal dose measurements are taken by a radiation detector in order to characterise the radiation beams from medical linear accelerators. Typically, an ionisation chamber and water phantom are used to create these radiation dose profiles. Water is used due to its tissue equivalence.
Transverse dose measurements are performed in the x (crossplane) or y (inplane) directions perpendicular to the radiation beam, and at a given depth (z) in the phantom. These are known dose profiles. Dose measurements taken along the z direction create radiation dose distribution known as a depth-dose curve.
See also
This article is issued from Wikipedia - version of the Saturday, March 16, 2013. The text is available under the Creative Commons Attribution/Share Alike but additional terms may apply for the media files.