Ion-mobility spectrometry
Ion-mobility spectrometry (IMS) is an analytical technique used to separate and identify ionized molecules in the gas phase based on their mobility in a carrier buffer gas. Though heavily employed for military or security purposes, such as detecting drugs and explosives, the technique also has many laboratory analytical applications, recently being coupled with mass spectrometry and high-performance liquid chromatography. IMS devices come in a wide range of sizes (often tailored for a specific application) and are capable of operating under a broad range of conditions. Systems operated at higher pressure (i.e. atmospheric conditions, 1 atm or 1013 mbar) are also accompanied by elevated temperature (above 100 °C), while lower pressure systems (1-20 mbar) do not require heating.
History
IMS was first developed primarily by Earl W. McDaniel of Georgia Institute of Technology in the 1950s and 1960s when he used drift cells with low applied electric fields to study gas phase ion mobilities and reactions.[1] In the following decades, he coupled his new technique with a magnetic-sector mass spectrometer, with others also utilizing his techniques in new ways. IMS cells have since been attached to many other mass spectrometers and high-performance liquid chromatography setups. Currently IMS is a widely used technique implemented by many, and improvements and other uses are continually being developed.
Outside of laboratory purposes, IMS has found great usage as a detection tool. More than 10,000 IMS devices are in use worldwide in airports, and the US Army has more than 50,000 IMS devices.[2] In industrial settings, uses of IMS include checking equipment cleanliness and detecting emission contents, such as determining the amount of hydrochloric and hydrofluoric acid in a stack gas from a process.[3]
Ion mobility
In the traditional method of drift-time IMS, commonly referred to as just IMS, produced ions travel through a drift tube which has an applied electric field and a carrier buffer gas that opposes the ion motion. At the end of the tube is a detector. Based on an ion’s mass, charge, size and shape (the ion mobility), the migration time through the tube is characteristic of different ions, leading to the ability to distinguish different analyte species. The area of an ion that gas molecules strike is an ion’s collision cross-section, related to the ion size and shape. The greater this collision cross-section is, meaning the larger the ion size, the more area available for buffer gas to collide and impede the ion’s drift – the ion then requires a longer time to migrate through the drift tube.
The physical quantity ion mobility K is defined as the proportionality factor of an ion's drift velocity vd in a gas and an electric field of strength E,
Ion mobilities are commonly reported as a reduced mobilities, correcting to standard gas density n0, which can be expressed in standard temperature T0 = 273 K and standard pressure p0 = 1013 mbar:
The ion mobility K can be experimentally determined by measuring the drift time tD of an ion traversing within a homogeneous electric field the potential difference U in the drift length L:
The ion mobility K can also be calculated by the Mason equation:
where Q is the ion charge, n is the drift gas number density, μ is the reduced mass of the ion and the drift gas molecules, k is Boltzmann constant, T is the drift gas temperature, and σ is the ion’s collision cross section with the drift gas. This relation holds approximately at a low electric field limit, where the ratio of E/N is small, at ≤ 6 x 10−17 J•C−1•cm2
A drift tube’s resolving power RP can be calculated as
where tD is the ion drift time, ΔtD is the Full width at half maximum, L is the tube length, E is the electric field strength, Q is the ion charge, k is Boltzmann’s constant, and T is the drift gas temperature.
With a low electric field applied, the thermal energy of the ions is greater than the energy gained from the electric field between collisions. With these ions having similar energies as the buffer gas molecules, diffusion forces dominate ion motion.
Instrumentation
Ionization
The molecules of the sample need to be ionized, usually by corona discharge, atmospheric pressure photoionization (APPI), electrospray ionization (ESI), or a radioactive source, e.g. a small piece of 63Ni or 241Am, similar to the one used in ionization smoke detectors. ESI and MALDI techniques are commonly used when IMS is paired with mass spec.
Doping materials are sometimes added to the drift gas for ionization selectivity. For example, acetone can be added for chemical warfare agent detection, chlorinated solvents added for explosives, and nicotinamide added for drugs detection.[4]
Analyzer
TOFIMS
In its simplest form an IMS system measures how fast a given ion moves in a uniform electric field through a given atmosphere.
In specified intervals, a sample of the ions is let into the drift chamber; the gating mechanism is based on a charged electrode working in a similar way as the control grid in triodes works for electrons. For precise control of the ion pulse width admitted to the drift tube, more complex gating systems such as a Bradbury-Nielsen design are employed. Once in the drift tube, ions are subjected to a homogeneous electric field ranging from a few volts per centimeter up to many hundreds of volts per centimeter. This electric field then drives the ions through the drift tube where they interact with the neutral drift molecules contained within the system.
In the drift tube, chemical species separate based on the ion mobility, arriving at the detector for measurement. Ions are recorded at the detector in order from the fastest to the slowest, generating a response signal characteristic for the chemical composition of the measured sample.
For drift-time IMS, two main methods are used – either reduced pressure or ambient pressure. A reduced pressure is where the applied pressure gas is at a few torr, commonly used to measure ion collision cross-sections. Ambient pressure is what is used for stand-alone detector devices, as well as detection for gas, liquid, and supercriticial fluid chromatographies. The higher pressures of ambient pressure methods allow for higher resolving power and greater separation selectivity due to a higher rate of ion-molecule interactions. Reduced pressure IMS allows for ion-focusing and an easier interface with mass spec.[1] Though drift electric fields are normally uniform, non-uniform drift fields are also used. These non-uniform field IMS results are often calibrated to uniform field instrumentation. Non-uniform field usage is still in a relatively early developing stage.[5]
DMS/FAIMS
DMS (differential mobility spectrometer) make use of the dependence of ion mobility K on electric field strength E at high electric fields. Ions are subjected to different field strengths for different amounts of time. Thereby only ions with certain mobility dependence survive. These types of IMS work as a scanable filter. They are also called FAIMS
DMA
DMA differential mobility analyzer make use of a fast gas stream perpendicular to the electric field. Thereby ions of different mobilities undergo different trajectories. This type of IMS corresponds to the sector instruments in mass spectrometry. They also work as a scanable filter. Examples include DMD (Differential Mobility Detector), first commercialized in Varian CP-4900 MicroGC.
Drift gas
The drift gas pressure is an important parameter for the IMS instrument design and resolution. Most drift gases have a higher break down voltage at pressures higher than a few mbar, with the break down voltage increasing as the pressure increases. As an example, high drift voltages of about 10-30 kV can be used with tubes of 1 m length and high gas pressures of 100-1000 mbar to obtain high resolutions. At higher pressures than 10 mbar, ions become more difficult to store. At lower pressures, ions can be stored more easily to obtain an accumulated continuous signal, with the trade-off of lower electric fields (around 10-30 V/cm). Elevated gas temperature assists in removing ion clusters that may distort experimental measurements.[6][7]
Detector
Often the detector is a simple Faraday plate, however, more advanced ion mobility instruments are coupled with mass spectrometers where both size and mass information may be obtained simultaneously.
Trapped ion mobility spectrometry
In trapped ion mobility spectrometry (TIMS) ions are held stationary (or trapped) in a flowing buffer gas by an axial electric field gradient (EFG) profile while the application of radio frequency (rf) potentials result in trapping in the radial dimension.[8] TIMS operates in the pressure range of 2 to 5 mbar and replaces the ion funnel found in the source region of modern mass spectrometers. TIMS can be coupled with nearly any mass analyzer through either the standard mode of operation for beam-type instruments, or selective accumulation mode (SAIMS) when used with trapping mass spectrometry (MS) instruments. TIMS devices do not require large size or high voltage in order to achieve high resolution, for instance achieving over 250 resolving power from a 4.7 cm device.[9] In addition TIMS is capable of higher sensitivity than traditional ion mobility systems because no grids or shutters exist in the ion path, improving ion transmission both during ion mobility experiments and while operating in a transparent MS only mode.
Combined IMS methods
IMS can be combined with other separation techniques.
GC-IMS : gas chromatography - ion mobility spectrometry
When IMS is coupled with gas chromatography, common sample introduction is with the GC capillary column directly connected to the IMS setup, with molecules ionized as they elute from GC.[4] A similar technique is commonly used for HPLC. A novel design for corona discharge ionization ion mobility spectrometry (CD–IMS) as a detector after capillary gas chromatography has been produced in 2012. In this design, a hollow needle was used for corona discharge creation and the effluent was entered into the ionization region on the upstream side of the corona source. In addition to the practical conveniences in coupling the capillary to IMS cell, this direct axial interfacing helps us to achieve a more efficient ionization, resulting in higher sensitivity.
IMS-MS : ion mobility spectrometry - mass spectrometry
When IMS is used with mass spectrometry, ion mobility spectrometry-mass spectrometry offers many advantages, including better signal to noise, isomer separation, and charge state identification.[1][10] IMS has commonly been attached to several mass spec analyzers, including quadropole, time-of-flight, and Fourier transform cyclotron resonance.
LC-IMS-MS : liquid chromatography - ion mobility spectrometry - mass spectrometry
Coupled with LC and MS, IMS has become widely used to analyze biomolecules, a practice heavily developed by David E. Clemmer, now at Indiana University (Bloomington).[5]
Applications
Perhaps ion mobility spectrometry's greatest strength is the speed at which separations occur—typically on the order of tens of milliseconds. This feature combined with its ease of use, relatively high sensitivity, and highly compact design have allowed IMS as a commercial product to be used as a routine tool for the field detection of explosives, drugs, and chemical weapons. Major manufacturers of IMS screening devices used in airports are Morpho and Smiths Detection.
In the pharmaceutical industry IMS is used in cleaning validations, demonstrating that reaction vessels are sufficiently clean to proceed with the next batch of pharmaceutical product. IMS is much faster and more accurate than HPLC and total organic carbon methods previously used. IMS is also used for analyzing the composition of drugs produced, thereby finding a place in quality assurance and control.[11] As a research tool ion mobility is becoming more widely used in the analysis of biological materials, specifically, proteomics and metabolomics. For example, IMS-MS using MALDI as the ionization method has helped make advances in proteomics, providing faster high-resolution separations of protein pieces in analysis.[12]
Aspirating IMS is an ion mobility spectrometry technology used to detect low or trace quantities of chemicals in the surrounding atmosphere. It is applied in industrial and military purposes to detect harmful substances in air.[13] Aspiration IMS operates with open-loop circulation of sampled air. Sample flow is passed via ionization chamber and then enters to measurement area where the ions are deflected into one or more measuring electrodes by perpendicular electric field which can be either static or varying. The output of the sensor is characteristic of the ion mobility distribution and can be used for detection and identification purposes.
In metabolomics the IMS is used to detect lung cancer, Chronic obstructive pulmonary disease, Sarcoidosis, potential rejections after Lung transplantation and relations to bacteria within the lung, see Breath gas analysis.
See also
References
- 1 2 3 Kanu AB, Dwivedi P, Tam M, Matz L, Hill HH (January 2008). "Ion mobility-mass spectrometry". J Mass Spectrom 43 (1): 1–22. doi:10.1002/jms.1383. PMID 18200615.
- ↑ Zolotov, Yu. A. (2006). “Ion Mobility Spectrometry.” Journal of Analytical Chemistry, 61 (6): 519.
- ↑ Particle Measuring Systems, Inc. “Ion Mobility Spectrometry (IMS) Theory and Applications”
- 1 2 Creaser, Colin; Thomas, Paul; et al. (2004). “Ion mobility spectrometry: a review. Part 1. Structural analysis by mobility measurement.” The Analyst, 129: 984-994.
- 1 2 Clemmer, David E. et al. (2008). “Biomolecule Analysis by Ion Mobility Spectrometry.” Annual Review of Analytical Chemistry, 1: 293-397.
- ↑ Bengt Nolting, Methods in Modern Biophysics, Springer Verlag, 2005, ISBN 3-540-27703-X
- ↑ Gary Eiceman & Zeev Karpas, Ion Mobility Spectrometry, CRC Press, 2005, ISBN 0-8493-2247-2
- ↑
- M. A. Park, Apparatus and Method for Parallel Flow Ion Mobility Spectrometry Combined with Mass Spectrometry, USPN 8,288,717
- ↑ Silveira, Joshua A.; Ridgeway, Mark E.; Park, Melvin A. (2014). "High Resolution Trapped Ion Mobility Spectrometery of Peptides". Analytical Chemistry: 140528090048009. doi:10.1021/ac501261h. ISSN 0003-2700.
- ↑ Fenn LS, McLean JA (June 2008). "Biomolecular structural separations by ion mobility-mass spectrometry". Anal Bioanal Chem 391 (3): 905–9. doi:10.1007/s00216-008-1951-x. PMID 18320175.
- ↑ O’Donnell, Ryan M.; Sun, Xiaobo; de B. Harrington, Peter. (2008). “Pharmaceutical applications of ion mobility spectrometry.” Trends in Analytical Chemistry, 27 (1): 44-53.
- ↑ McLean, J.A. . et al. (2005). “Ion mobility–mass spectrometry: a new paradigm for proteomics.” International Journal of Mass Spectrometry, 240: 301-315.
- ↑ Riikka-Marjaana Räsänen, Marjaana Nousiainen, Kaleva Peräkorpi, Mika Sillanpää, Lauri Polari, Osmo Anttalainen and Mikko Utriainen. "Analytica Chimica Acta", Volume 623, Issue 1, 8 August 2008, Pages 59-65
Bibliography
- G.A. Eiceman, Z. Karpas (2005). Ion Mobility Spectrometry (2nd ed.). Boca Raton, FL., USA: CRC Press. ISBN 9780849322471.
- Alexandre A. Shvartsburg (2008). Differential Ion Mobility Spectrometry: Nonlinear Ion Transport and Fundamentals of FAIMS (2nd ed.). Boca Raton, FL., USA: CRC Press. ISBN 9781420051063.
- Charles L. Wilkins, Sarah Trimpin, ed. (2010). Ion Mobility Spectrometry - Mass Spectrometry: Theory and Applications. Boca Raton, FL., USA: CRC Press. ISBN 9781439813249.
- Joachim Stach, Jörg I. Baumbach: Ion Mobility Spectrometry – Basic Elements and Applications. In: International Journal for Ion Mobility Spectrometry. Vol. 5, Nr. 1, 2002, ISSN 1435-6163, S. 1–21, Digitalisat (PDF; 1,1 MB).
External links
Listing of analytical instrument sites at DMOZ