Light sheet fluorescence microscopy

The principle setup of a light sheet fluorescence microscope.

Light sheet fluorescence microscopy (LSFM) is a fluorescence microscopy technique with an intermediate optical resolution, but good optical sectioning capabilities and high speed. In contrast to epifluorescence microscopy only a thin slice (usually a few hundred nanometers to a few micrometers) of the sample is illuminated perpendicularly to the direction of observation. For illumination, a laser light-sheet is used, i.e. a laser beam which is focused only in one direction (e.g. using a cylindrical lens). A second method uses a circular beam scanned in one direction to create the lightsheet. As only the actually observed section is illuminated, this method reduces the photodamage and stress induced on a living sample. Also the good optical sectioning capability reduces the background signal and thus creates images with higher contrast, comparable to confocal microscopy. Because LSFM scans samples by using a plane of light instead of a point (as in confocal microscopy), it can acquire images at speeds 100 to 1000 times faster than those offered by point-scanning methods.

Comparison of different microscopy illumination modalities (LSFM: lightsheet fluorescence microscopy, WF: widefield microscopy, CF: confocal microscopy). LSFM combines good z-sectioning (as confocal) and only illuminates the observed plane

This method is used in cell biology[1] and for microscopy of whole living creatures, such as embryos.[2] It is now often used for the longtime observation of embryonal development in different model organisms.

Starting in 1994, LSFM was developed as orthogonal plane fluorescence optical sectioning microscopy or tomography (OPFOS)[3] mainly for large samples and later as the selective/single plane illumination microscopy (SPIM) also with sub-cellular resolution.[4] This introduced an illumination scheme into fluorescence microscopy, which has already been used successfully for dark field microscopy under the name ultramicroscopy.[5]

Setup of LSFM

Basic setup

Illustration of different LSFM implementations. See text for details. Legend: CAM=camera, TL=tube lens, F=filter, DO=detection objective, S=sample, SC=sample chamber, PO=projection objective, CL=cylindrical lens, SM=scanning mirror

In this type of microscopy,[6] the illumination is done perpendicularly to the direction of observation (see schematic image at the top of the article). The expanded beam of a laser is focused in only one direction by a cylindrical lens, or by a combination of a cylindrical lens and a microscope objective as the latter is available in better optical quality and with higher numerical aperture than the first. This way a thin sheet of light or lightsheet is created in the focal region that can be used to excite fluorescence only in a thin slice (usually a few micrometers thin) of the sample.

The fluorescence light emitted from the lightsheet is then collected perpendicularly with a standard microscope objective and projected onto an imaging sensor (usually a CCD, electron multiplying CCD or CMOS camera). In order to let enough space for the excitation optics/lightsheet an observation objective with high working distance is used. In most LSFMs the detection objective and sometimes also the excitation objective are fully immersed in the sample buffer, so usually the sample and excitation/detection optics are embedded into a buffer-filled sample chamber, which can also be used to control the environmental conditions (temperature, carbon dioxide level ...) during the measurement. The sample mounting in LSFM is described below in more detail.

As both the excitation lightsheet and the focal plane of the detection optics have to coincide to form an image, focussing different parts of the sample can not be done by translating the detection objective, but usually the whole sample is translated and rotated instead.

Extensions of the basic LSFM idea

In recent years, several extensions to this scheme have been developed:

Sample mounting

Different types of sample mounting for LSFM: embryo embedded in hanging gel cylinder, plant growing in supported gel cylinder, adherent cells on glass, liquid sample in a sample bag

The separation of the illumination and detection beampaths in LSFM (except in oblique plane microscopy) creates a need for specialized sample mounting methods. To date most LSFMs are built in such a way that the illumination and detection beampath lie in a horizontal plane (see illustrations above), thus the sample is usually hanging from the top into the sample chamber or is resting on a vertical support inside the sample chamber. Several methods have been developed to mount all sorts of samples:

Some LSFMs have been developed where the sample is mounted as in standard microscopy (e.g. cells grow horizontally on the bottom of a petri dish) and the excitation and detection optics are constructed in an upright plane from above. This also allows combining a LSFM with a standard inverted microscope and avoids the requirement for specialized sample mounting procedures.[15][22][23][24]

LSFM image properties

Typical imaging modes

Most LSFMs are used to produce 3D images of the sample by moving the sample through the image plane. If the sample is larger than the field of view of the image sensor, the sample also has to be shifted laterally. An alternative approach is to move the image plane through the sample to create the image stack.[24]

Long experiments can be carried out, for example with stacks is recorded every 10 s–10 min over the timespan of days. This allows study of changes over time in 3D, or so-called 4D microscopy.

After the image acquisition the different image stacks are registered to form one single 3D dataset. Multiple views of the sample can be collected, either by interchanging the roles of the objectives [24] or by rotating the sample.[7] Having multiple views can yield more information than a single stack; for example occlusion of some parts of the sample may be overcome. Multiple views also improves 3D image resolution by overcoming poor axial resolution as described below.

Some studies also use a SPIM to image only one slice of the sample, but at much higher temporal resolution. This allows e.g. to observe the beating heart of a zebra fish embryo in real-time.[25] Together with fast translation stages for the sample a high-speed 3D particle tracking has been implemented.[26]

Power of resolution

The lateral resolution of a SPIM is comparable to that of a standard (epi) fluorescence microscope, as it is determined fully by the detection objective and the wavelength of the detected light (see Abbe limit). E.g. for detection in the green spectral region around 525 nm, a resolution of 250–500 nm can be reached.[6] The axial resolution is worse than the lateral (about a factor of 4), but it can be improved by using a thinner lightsheet in which case nearly isotropic resolution is possible.[15] Thinner light sheets are either thin only in a small region (for Gaussian beams) or else specialized beam profiles such as Bessel beams must be used (besides added complexity, such schemes add side lobes which can be detrimental [12]). Alternatively, isotropic resolution can be achieved by computationally combining 3D image stacks taken from the same sample under different angles. Then the depth-resolution information lacking in one stack is supplied from another stack; for example with two orthogonal stacks the (poor-resolution) axial direction in one stack is a (high-resolution) lateral direction in the other stack.

The lateral resolution of LSFM can be improved beyond the Abbe limit, by using super resolution microscopy techniques, e.g. with using the fact, that single fluorophores can be located with much higher spatial precision than the nominal resolution of the used optical system (see stochastic localization microscopy techniques).[18] Also structured illumination techniques have been applied to further improve the optical sectioning capacity of LSFM.[19]

Stripe artifacts

Top: stripe artifacts in LSFM. Bottom: Reducing stripe artifacts by pivoting

As the illumination typically penetrates the sample from one side, obstacles lying in the way of the lightsheet can disturb its quality by scattering and/or absorbing the light. This typically leads to dark and bright stripes in the images. If parts of the samples have a significantly higher refractive index (e.g. lipid vesicles in cells), they can also lead to a focussing effect resulting in bright stripes behind these structures. To overcome this artifact, the lightsheets can e.g. be "pivoting". That means that the lightsheet's direction of incidence is changed rapidly (~1 kHz rate) by a few degrees (~10°), so light also hits the regions behind the obstacles. In addition illumination can be performed with two (pivoted) lightsheets (see above) to further reduce these artifacts.[7]

History

At the beginning of the 20th century, R. A. Zsigmondy introduced the ultramicroscope as a new illumination scheme into dark-field microscopy. Here sunlight or a white lamp is used to illuminate a precision slit. The slit is then imaged by a condensor lens into the sample to form a lightsheet. Scattering (sub-diffractive) particles can be observed perpendicularly with a microscope. This setup allowed the observation of particles with sizes smaller than the microscope's resolution and led to a Nobel prize for Zsigmondy in 1925.[27]

The first application of this illumination scheme for fluorescence microscopy was published in 1993 by Voie et al. under the name orthogonal-plane fluorescence optical sectioning (OPFOS).[3] for imaging of the internal structure of the cochlea. The resolution at that time was limited to 10 µm laterally and 26 µm longitudinally but at a sample size in the millimeter range. The OPFOS microscope used a simple cylindrical lens for illumination. Further development and improvement of the SPIM started in 2004.[4] After this publication by Huisken et al. the technique found wide application and is still adapted to new measurement situations today (see above). Since 2010 a first ultramicroscope with fluorescence excitation and limited resolution[28] and since 2012 a first SPIM are available commercially.[29] A good overview about the development of SPIM is given in ref.[30] During 2012 also open source projects have started to appear that freely publish complete construction plans for LSFMs and also the required software suites.[31][32][33][34]

Application

SPIM/LSFM is often used in developmental biology, where it enables long-time (several days) observations of embryonal development (even with full lineage tree reconstruction).[4][35] SPIM can also be combined with techniques, like fluorescence correlation spectroscopy to allow spatially resolved mobility measurements of fluorescing particles (e.g. fluorescent beads, quantum dots or fluorescently labeled proteins) inside living biological samples.[15][16]

  1. ^ {{cite journal| author = Corinne Lorenzo
  2. ^ Daisuke Takao, Atsushi Taniguchi, Takaaki Takeda, Seiji Sonobe, Shigenori Nonaka, Alexandre J. Kabla (2012-12-05), [[1] "High-Speed Imaging of Amoeboid Movements Using Light-Sheet Microscopy"] (in German), PLoS ONE 7 (12): pp. e50846, doi:10.1371/journal.pone.0050846, ISSN 1932-6203, [2] 

References

  1. Philipp J. Keller, Ernst H. K. Stelzer: Lichtscheiben-Mikroskopie in der molekularen Zellbiophysik In: LABORWELT. 7. Jahrgang, Nr. 5, 2006, S. 18–21 (Online-Version).
  2. U. Krzic, S. Günther, L. Hufnagel, D. von Gegerfelt, H. Karlsson, E. Illy, J. Hel: Lichtscheiben-Fluoreszenzmikroskopie (SPIM) und Laser-Anregung in orange zur Abbildung lebender Organismen. In: BioPhotonik. Nr. 1, 2011, S. 42–44 (Online-Version).
  3. 1 2 A. H. Voie, D. H. Burns, F. A. Spelman (June 1993). "Orthogonal-plane fluorescence optical sectioning: Three-dimensional imaging of macroscopic biological specimens". Journal of Microscopy 170 (3): 229–236. doi:10.1111/j.1365-2818.1993.tb03346.x. ISSN 0022-2720.
  4. 1 2 3 Huisken, J.; Swoger, J.; Del Bene, F.; Wittbrodt, J.; Stelzer, E. H. (2004). "Optical sectioning deep inside live embryos by selective plane illumination microscopy". Science 305 (5686): 1007–1009. doi:10.1126/science.1100035. PMID 15310904.
  5. Timo Mappes, Norbert Jahr, Andrea Csaki, Nadine Vogler, Juergen Popp, Wolfgang Fritzsche (5 November 2012). "The Invention of Immersion Ultramicroscopy in 1912-The Birth of Nanotechnology?". Angewandte Chemie International Edition 51 (45): 11208–11212. doi:10.1002/anie.201204688. ISSN 1433-7851.
  6. 1 2 K. Greger, J. Swoger, E. H. Stelzer: Basic building units and properties of a fluorescence single plane illumination microscope. In: The Review of scientific instruments. Band 78, Nr. 2, 2007, ISSN 0034-6748, S. 023705, PMID 17578115.
  7. 1 2 3 Huisken, Jan; Stainier, Didier Y. R. (2007). "Even fluorescence excitation by multidirectional selective plane illumination microscopy (mSPIM)". Optics Letters 32 (17): 2608–10. doi:10.1364/OL.32.002608. ISSN 0146-9592. PMID 17767321. Retrieved 2012-10-27.(subscription required)
  8. Tomer, Raju; Khairy, Khaled; Amat, Fernando; Keller, Philipp J (3 June 2012). "Quantitative high-speed imaging of entire developing embryos with simultaneous multiview light-sheet microscopy". Nature Methods 9 (7): 755–763. doi:10.1038/nmeth.2062. ISSN 1548-7091. PMID 22660741. Retrieved 2012-10-27.(subscription required)
  9. Krzic, Uros; Gunther, Stefan; Saunders, Timothy E; Streichan, Sebastian J; Hufnagel, Lars (3 June 2012). "Multiview light-sheet microscope for rapid in toto imaging". Nature Methods 9 (7): 730–733. doi:10.1038/nmeth.2064. ISSN 1548-7091. PMID 22660739. Retrieved 2012-10-27.(subscription required)
  10. Keller, P. J.; Schmidt, A. D.; Wittbrodt, J.; Stelzer, E.H.K. (14 November 2008). "Reconstruction of Zebrafish Early Embryonic Development by Scanned Light Sheet Microscopy" (pdf). Science 322 (5904): 1065–1069. doi:10.1126/science.1162493. ISSN 0036-8075. PMID 18845710. Retrieved 2012-10-27.
  11. F. O. Fahrbach, A. Rohrbach: A line scanned light-sheet microscope with phase shaped self-reconstructing beams. In: Optics express. Band 18, Nummer 23, November 2010, S. 24229–24244, ISSN 1094-4087. PMID 21164769.
  12. 1 2 Planchon, T. A.; Gao, L.; Milkie, D. E.; Davidson, M. W.; Galbraith, J. A.; Galbraith, C. G.; Betzig, E. (2011). "Rapid three-dimensional isotropic imaging of living cells using Bessel beam plane illumination". Nature methods. 8 (5): 417–423. doi:10.1038/nmeth.1586. PMID 21378978.
  13. Dunsby, C. (2008). "Optically sectioned imaging by oblique plane microscopy". Optics Express 16 (25): 20306. doi:10.1364/OE.16.020306. ISSN 1094-4087. Retrieved 2012-10-27.
  14. Zeno Lavagnino; Francesca Cella Zanacchi; Emiliano Ronzitti; Alberto Diaspro (2013). "Two-photon excitation selective plane illumination microscopy (2PE-SPIM) of highly scattering samples: characterization and application". Optics Express 21 (5): 5998–6008. doi:10.1364/OE.21.005998. ISSN 1094-4087. PMID 23482168.
  15. 1 2 3 4 Capoulade, J.; Wachsmuth, M.; Hufnagel, L.; Knop, M. (2011). "Quantitative fluorescence imaging of protein diffusion and interaction in living cells". Nature biotechnology. 29 (9): 835–839. doi:10.1038/nbt.1928. PMID 21822256.
  16. 1 2 3 T. Wohland, X. Shi, J. Sankaran, E. H. Stelzer: Single plane illumination fluorescence correlation spectroscopy (SPIM-FCS) probes inhomogeneous three-dimensional environments. In: Optics express. Band 18, Nr. 10, ISSN 1094-4087, 2010, S. 10627–10641, PMID 20588915.
  17. Klaus Greger, Manuel J. Neetz, Emmanuel G. Reynaud, Ernst H.K. Stelzer (2011). "Three-dimensional Fluorescence Lifetime Imaging with a Single Plane Illumination Microscope provides an improved Signal to Noise Ratio". Optics Express 19 (21): 20743–50. doi:10.1364/OE.19.020743. ISSN 1094-4087. PMID 21997084.
  18. 1 2 Francesca Cella Zanacchi, Zeno Lavagnino, Michela Perrone Donnorso, Alessio Del Bue, Laura Furia, Mario Faretta, Alberto Diaspro (9 October 2011). "Live-cell 3D super-resolution imaging in thick biological samples". Nature Methods 8 (12): 1047–1049. doi:10.1038/nmeth.1744. ISSN 1548-7091. PMID 21983925.
  19. 1 2 Jerome Mertz; Jinhyun Kim (2010). "Scanning light-sheet microscopy in the whole mouse brain with HiLo background rejection". Journal of Biomedical Optics 15 (1): 016027. doi:10.1117/1.3324890. ISSN 1083-3668. PMC 2917465. PMID 20210471.
  20. Friedrich, Mike; Gan, Qiang; Ermolayev, Vladimir; Harms, Gregory S. (2011). "STED-SPIM: Stimulated Emission Depletion Improves Sheet Illumination Microscopy Resolution". Biophysical Journal 100 (8): L43–5. doi:10.1016/j.bpj.2010.12.3748. PMC 3077687. PMID 21504720..
  21. Alexis Maizel, Daniel von Wangenheim, Fern n Federici, Jim Haseloff, Ernst H.K. Stelzer (October 2011). "High-resolution live imaging of plant growth in near physiological bright conditions using light sheet fluorescence microscopy". The Plant Journal 68 (2): 377–385. doi:10.1111/j.1365-313X.2011.04692.x. ISSN 0960-7412. PMID 21711399.
  22. Terrence F. Holekamp, Diwakar Turaga, Timothy E. Holy (13 March 2008). "Fast Three-Dimensional Fluorescence Imaging of Activity in Neural Populations by Objective-Coupled Planar Illumination Microscopy". Neuron 57 (5): 661–672. doi:10.1016/j.neuron.2008.01.011. ISSN 0896-6273. PMID 18341987.
  23. Y. Wu, A. Ghitani, R. Christensen, A. Santella, Z. Du, G. Rondeau, Z. Bao, D. Colon-Ramos, H. Shroff (25 October 2011). "Inverted selective plane illumination microscopy (iSPIM) enables coupled cell identity lineaging and neurodevelopmental imaging in Caenorhabditis elegans". Proceedings of the National Academy of Sciences 108 (43): 17708–17713. doi:10.1073/pnas.1108494108. ISSN 0027-8424.
  24. 1 2 3 Wu, Yicong; Wawrzusin, Peter; Senseney, Justin; Fischer, Robert S; Christensen, Ryan; Santella, Anthony; York, Andrew G; Winter, Peter W; Waterman, Clare M; Bao, Zhirong; Colón-Ramos, Daniel A; McAuliffe, Matthew; Shroff, Hari (2013). "Spatially isotropic four-dimensional imaging with dual-view plane illumination microscopy". Nature Biotechnology 31 (11): 1032–1038. doi:10.1038/nbt.2713. ISSN 1087-0156.
  25. "Webpage Huisken Lab".
  26. Jörg G. Ritter, Roman Veith, Jan-Peter Siebrasse, Ulrich Kubitscheck (2008), [ "High-contrast single-particle tracking by selective focal plane illumination microscopy"] (in German), Optics Express 16 (10): pp. 7142, doi:10.1364/OE.16.007142, ISSN 1094-4087,
  27. Nobel lecture of R. A. Zsigmondy: Properties of colloids (including a short explanation of the ultramicroscope)
  28. press release of LaVision Biotech (accessed 2012-11-04)
  29. Carl Zeiss press release about the Lightsheet Z.1 Light Sheet Microscope System (accessed 2012-11-15)
  30. P. A. Santi (1 February 2011). "Light Sheet Fluorescence Microscopy: A Review". Journal of Histochemistry & Cytochemistry 59 (2): 129–138. doi:10.1369/0022155410394857. ISSN 0022-1554.
  31. OpenSPIM project webpage (accessed 2013-06-08)
  32. Peter G Pitrone, Johannes Schindelin, Luke Stuyvenberg, Stephan Preibisch, Michael Weber, Kevin W Eliceiri, Jan Huisken, Pavel Tomancak (9 June 2013). "OpenSPIM: an open-access light-sheet microscopy platform". Nature Methods 10 (7): 598–9. doi:10.1038/nmeth.2507. ISSN 1548-7091. PMID 23749304.
  33. The OpenSPIN project webpage (accessed 2013-06-08)
  34. Emilio J Gualda, Tiago Vale, Pedro Almada, Jos A Feij , Gabriel G Martins, Nuno Moreno (9 June 2013). "OpenSpinMicroscopy: an open-source integrated microscopy platform". Nature Methods 10 (7): 599–600. doi:10.1038/nmeth.2508. ISSN 1548-7091. PMID 23749300.
  35. Verveer, P. J.; Swoger, J.; Pampaloni, F.; Greger, K.; Marcello, M.; Stelzer, E. H. (2007). "High-resolution three-dimensional imaging of large specimens with light sheet-based microscopy". Nature methods 4 (4): 311–313. doi:10.1038/nmeth1017. PMID 17339847.