Prothrombin time
Prothrombin time | |
---|---|
Intervention | |
Blood plasma after the addition of tissue factor. The gel-like structure is strong enough to hold a steel ball. | |
MeSH | D011517 |
The prothrombin time (PT) — along with its derived measures of prothrombin ratio (PR) and international normalized ratio (INR) — are assays evaluating the extrinsic pathway of coagulation. This test is also called "ProTime INR" and "PT/INR". They are used to determine the clotting tendency of blood, in the measure of warfarin dosage, liver damage, and vitamin K status. PT measures factors I (fibrinogen), II (prothrombin), V, VII, and X. It is used in conjunction with the activated partial thromboplastin time (aPTT) which measures the intrinsic pathway and common pathway.
Laboratory measurement
The reference range for prothrombin time depends on the analytical method used, but is usually around 12–13 seconds (results should always be interpreted using the reference range from the laboratory that performed the test), and the INR in absence of anticoagulation therapy is 0.8-1.2. The target range for INR in anticoagulant use (e.g. warfarin) is 2 to 3. In some cases, if more intense anticoagulation is thought to be required, the target range may be as high as 2.5-3.5 depending on the indication for anticoagulation.[1] In The Netherlands, the target INR for 'low intensity' is between 2.5 and 3.5 and for 'high intensity' between 3.0 and 4.0.[2]
Methodology
Prothrombin time is typically analyzed by a laboratory technologist on an automated instrument at 37 °C (as a nominal approximation of normal human body temperature).
- Blood is drawn into a test tube containing liquid sodium citrate, which acts as an anticoagulant by binding the calcium in a sample. The blood is mixed, then centrifuged to separate blood cells from plasma (as prothrombin time is most commonly measured using blood plasma). In newborns, a capillary whole blood specimen is used.[3]
- A sample of the plasma is extracted from the test tube and placed into a measuring test tube (Note: for an accurate measurement, the ratio of blood to citrate needs to be fixed and should be labeled on the side of the measuring test tube by the manufacturing company; many laboratories will not perform the assay if the tube is underfilled and contains a relatively high concentration of citrate—the standardized dilution of 1 part anticoagulant to 9 parts whole blood is no longer valid).
- Next an excess of calcium (in a phospholipid suspension) is added to the test tube, thereby reversing the effects of citrate and enabling the blood to clot again.
- Finally, in order to activate the extrinsic / tissue factor clotting cascade pathway, tissue factor (also known as factor III) is added and the time the sample takes to clot is measured optically. Some laboratories use a mechanical measurement, which eliminates interferences from lipemic and icteric samples. The prothrombin ratio (aka international normalized ratio) is the prothrombin time for a patient sample divided by the result for control plasma.
International normalized ratio
The result (in seconds) for a prothrombin time performed on a normal individual will vary according to the type of analytical system employed. This is due to the variations between different batches of manufacturer's tissue factor used in the reagent to perform the test. The INR was devised to standardize the results. Each manufacturer assigns an ISI value (International Sensitivity Index) for any tissue factor they manufacture. The ISI value indicates how a particular batch of tissue factor compares to an international reference tissue factor. The ISI is usually between 1.0 and 2.0. The INR is the ratio of a patient's prothrombin time to a normal (control) sample, raised to the power of the ISI value for the analytical system being used.
Interpretation
The prothrombin time is the time it takes plasma to clot after addition of tissue factor (obtained from animals such as rabbits, or recombinant tissue factor, or from brains of autopsy patients). This measures the quality of the extrinsic pathway (as well as the common pathway) of coagulation. The speed of the extrinsic pathway is greatly affected by levels of functional factor VII in the body. Factor VII has a short half-life and the carboxylation of its glutamate residues requires vitamin K. The prothrombin time can be prolonged as a result of deficiencies in vitamin K, warfarin therapy, malabsorption, or lack of intestinal colonization by bacteria (such as in newborns). In addition, poor factor VII synthesis (due to liver disease) or increased consumption (in disseminated intravascular coagulation) may prolong the PT.
The INR is typically used to monitor patients on warfarin or related oral anticoagulant therapy. The normal range for a healthy person not using warfarin is 0.8–1.2, and for people on warfarin therapy an INR of 2.0–3.0 is usually targeted, although the target INR may be higher in particular situations, such as for those with a mechanical heart valve. If the INR is outside the target range, a high INR indicates a higher risk of bleeding, while a low INR suggests a higher risk of developing a clot.
Condition | Prothrombin time | Partial thromboplastin time | Bleeding time | Platelet count |
---|---|---|---|---|
Vitamin K deficiency or warfarin | Prolonged | Normal or mildly prolonged | Unaffected | Unaffected |
Disseminated intravascular coagulation | Prolonged | Prolonged | Prolonged | Decreased |
Von Willebrand disease | Unaffected | Prolonged or unaffected | Prolonged | Unaffected |
Haemophilia | Unaffected | Prolonged | Unaffected | Unaffected |
Aspirin | Unaffected | Unaffected | Prolonged | Unaffected |
Thrombocytopenia | Unaffected | Unaffected | Prolonged | Decreased |
Liver failure, early | Prolonged | Unaffected | Unaffected | Unaffected |
Liver failure, end-stage | Prolonged | Prolonged | Prolonged | Decreased |
Uremia | Unaffected | Unaffected | Prolonged | Unaffected |
Congenital afibrinogenemia | Prolonged | Prolonged | Prolonged | Unaffected |
Factor V deficiency | Prolonged | Prolonged | Unaffected | Unaffected |
Factor X deficiency as seen in amyloid purpura | Prolonged | Prolonged | Unaffected | Unaffected |
Glanzmann's thrombasthenia | Unaffected | Unaffected | Prolonged | Unaffected |
Bernard-Soulier syndrome | Unaffected | Unaffected | Prolonged | Decreased or unaffected |
Factor XII deficiency | Unaffected | Prolonged | Unaffected | Unaffected |
C1INH deficiency | Unaffected | Shortened | Unaffected | Unaffected |
Factors determining accuracy
Lupus anticoagulant, a circulating inhibitor predisposing for thrombosis, may skew PT results, depending on the assay used.[4] Variations between various thromboplastin preparations have in the past led to decreased accuracy of INR readings, and a 2005 study suggested that despite international calibration efforts (by INR) there were still statistically significant differences between various kits,[5] casting doubt on the long-term tenability of PT/INR as a measure for anticoagulant therapy.[6]
Statistics
An estimated 800 million PT/INR assays are performed annually worldwide.[6]
Near-patient testing
In addition to the laboratory method outlined above, near-patient testing (NPT) or home INR monitoring is becoming increasingly common in some countries. In the United Kingdom, for example, near-patient testing is used both by patients at home, and by some anticoagulation clinics (often hospital-based) as a fast and convenient alternative to the lab method. After a period of doubt about the accuracy of NPT results, a new generation of machines and reagents seems to be gaining acceptance for its ability to deliver results close in accuracy to those of the lab.[7]
In a typical NPT set up, a small table-top device is used. A drop of capillary blood is obtained with an automated finger-prick, which is almost painless. This drop is placed on a disposable test strip with which the machine has been prepared. The resulting INR comes up on the display a few seconds later. A similar form of testing is used by people with diabetes for monitoring blood sugar levels, which is easily taught and routinely practiced.
Local policy determines whether the patient or a coagulation specialist (pharmacist, nurse, general practitioner or hospital doctor) interprets the result and determines the dose of medication. In Germany and Austria, patients may adjust the medication dose themselves, while in the UK and the USA this remains in the hands of a health care professional.
A significant advantage of home testing is the evidence that patient self-testing with medical support and patient self-management (where patients adjust their own anticoagulant dose) improves anticoagulant control. A meta analysis which reviewed 14 trials showed that home testing led to a reduced incidence of complications (bleeding and thrombosis) and improved the time in the therapeutic range, which is an indirect measure of anticoagulant control.[8]
Other advantages of the NPT approach are that it is fast and convenient, usually less painful, and offers, in home use, the ability for patients to measure their own INRs when required. Among its problems are that quite a steady hand is needed to deliver the blood to the exact spot, that some patients find the finger-pricking difficult, and that the cost of the test strips must also be taken into account. In the UK these are available on prescription so that elderly and unwaged people will not pay for them and others will pay only a standard prescription charge, which at the moment represents only about 20% of the retail price of the strips. In the USA, NPT in the home is currently reimbursed by Medicare for patients with mechanical heart valves, while private insurers may cover for other indications. Medicare is now covering home testing for patients with chronic atrial fibrillation. Home testing requires a doctor's prescription and that the meter and supplies are obtained from a Medicare approved Independent Diagnostic Testing Facility (IDTF).
There is some evidence to suggest that NPT may be less accurate for certain patients, for example those who have the lupus anticoagulant.[9]
Guidelines
International guidelines were published in 2005 to govern home monitoring of oral anticoagulation by the International Self-Monitoring Association for Oral Anticoagulation.[10] The international guidelines study stated, "The consensus agrees that patient self-testing and patient self-management are effective methods of monitoring oral anticoagulation therapy, providing outcomes at least as good as, and possibly better than, those achieved with an anticoagulation clinic. All patients must be appropriately selected and trained. Currently available self-testing/self-management devices give INR results which are comparable with those obtained in laboratory testing."
Medicare coverage for home testing of INR has been expanded in order to allow more people access to home testing of INR in the USA. The release on 19 March 2008 said, "[t]he Centers for Medicare & Medicaid Services (CMS) expanded Medicare coverage for home blood testing of prothrombin time (PT) International Normalized Ratio (INR) to include beneficiaries who are using the drug warfarin, an anticoagulant (blood thinner) medication, for chronic atrial fibrillation or venous thromboembolism." In addition, "[t]hose Medicare beneficiaries and their physicians managing conditions related to chronic atrial fibrillation or venous thromboembolism will benefit greatly through the use of the home test."[11]
History
The prothrombin time was developed by Dr Armand Quick and colleagues in 1935,[12] and a second method was published by Dr Paul Owren,[13] also called the "p and p" or "prothrombin and proconvertin" method. It aided in the identification of the anticoagulants dicumarol and warfarin,[14] and was used subsequently as a measure of activity for warfarin when used therapeutically.
The INR was invented in the early 1980s by Tom Kirkwood working at the UK National Institute for Biological Standards and Control (and subsequently at the UK National Institute for Medical Research) to provide a consistent way of expressing the prothrombin time ratio, which had previously suffered from a large degree of variation between centres using different reagents. The INR was coupled to Dr Kirkwood's simultaneous invention of the International Sensitivity Index (ISI), which provided the means to calibrate different batches of thromboplastins to an international standard.[15] The INR became widely accepted worldwide, especially after endorsement by the World Health Organization.[16]
See also
- Partial thromboplastin time (PTT), or activated partial thromboplastin time (aPTT or APTT)
- Thrombin time (TT)
- Thromboelastography
- Thrombodynamics test
- D-dimer
- Thrombus
References
- ↑ "Warfarin Therapy Management in Adults" (PDF).
- ↑ De Kunst van het Doseren [The Art of Dosing] (PDF) (in Dutch). Federatie van Nederlandse Trombosediensten. February 2014. Retrieved July 1, 2014.
- ↑ Fritsma, George A. (2002). "Evaluation of Hemostasis." Hematology: Clinical Principles and Applications . Ed. Bernadette Rodak. W.B. Saunders Company: Philadelphia, 2002. 719-53. Print
- ↑ Della Valle P, Crippa L, Garlando AM, et al. (December 1999). "Interference of lupus anticoagulants in prothrombin time assays: implications for selection of adequate methods to optimize the management of thrombosis in the antiphospholipid-antibody syndrome" (PDF). Haematologica 84 (12): 1065–74. PMID 10586206.
- ↑ Horsti J, Uppa H, Vilpo JA (March 2005). "Poor agreement among prothrombin time international normalized ratio methods: comparison of seven commercial reagents". Clin. Chem. 51 (3): 553–60. doi:10.1373/clinchem.2004.043836. PMID 15665046.
- 1 2 Jackson CM, Esnouf MP (March 2005). "Has the time arrived to replace the quick prothrombin time test for monitoring oral anticoagulant therapy?". Clin. Chem. 51 (3): 483–5. doi:10.1373/clinchem.2004.045393. PMID 15738512.
- ↑ Poller L, Keown M, Chauhan N, et al. (September 2003). "European Concerted Action on Anticoagulation. Correction of displayed international normalized ratio on two point-of-care test whole-blood prothrombin time monitors (CoaguChek Mini and TAS PT-NC) by independent international sensitivity index calibration". Br. J. Haematol. 122 (6): 944–9. doi:10.1046/j.1365-2141.2003.04521.x. PMID 12956765.
- ↑ Heneghan C, Alonso-Coello P, Garcia-Alamino JM, Perera R, Meats E, Glasziou P (February 2006). "Self-monitoring of oral anticoagulation: a systematic review and meta-analysis". Lancet 367 (9508): 404–11. doi:10.1016/S0140-6736(06)68139-7. PMID 16458764.
- ↑ Moll, S and Ortel, TL. (August 1997). "Metering Warfarin Therapy in Patients with Lupus Anticoagulants.". Annals of Internal Medicine 127 (3): 177–185. doi:10.7326/0003-4819-127-3-199708010-00001. PMID 9245222.
- ↑ Jack Ansell (10 March 2005). "Guidelines for implementation of patient self-testing and patient self-management of oral anticoagulation. International consensus guidelines prepared by International Self-Monitoring Association for Oral Anticoagulation". International Journal of Cardiology 99 (1): 37–45. doi:10.1016/j.ijcard.2003.11.008. PMID 15721497.
- ↑ "Medicare expands coverage for home blood testing of prothrombin time international normalized ratio". The Centers for Medicare and Medicaid Services. 19 March 2008.
- ↑ Quick AJ, Stanley-Brown M, Bancroft FW (1935). "A study of the coagulation defect in hemophilia and in jaundice". Am J Med Sci 190 (4): 501. doi:10.1097/00000441-193510000-00009.
- ↑ Owren PA, Aas K (1951). "The control of dicumarol therapy and the quantitative determination of prothrombin and proconvertin". Scand. J. Clin. Lab. Invest. 3 (3): 201–8. doi:10.3109/00365515109060600. PMID 14900966.
- ↑ Campbell HA, Smith WK, Roberts WL, Link KP (1941). "Studies on the hemorrhagic sweet clover disease. II. The bioassay of hemorrhagic concentrates by following the prothrombin level in the plasma of rabbit blood". J Biol Chem 138: 1–20.
- ↑ Kirkwood TB (June 1983). "Calibration of reference thromboplastins and standardisation of the prothrombin time ratio". Thrombosis and Haemostasis 49 (3): 238–44. PMID 6879511.
- ↑ Anonymous (1983). "33: Expert Committee on Biological Standardization. Requirements for thromboplastins and plasma used to control oral anticoagulant therapy". World Health Organ Tech Rep Ser. pp. 81–105.
External links
- PT and INR - Lab Tests Online
|