Surface plasmon resonance microscopy

Surface plasmon resonance microscopy (SPRM), also called surface plasmon resonance imaging (SPRI),[1] is a label free analytical tool that combines the surface plasmon resonance of metallic surfaces with imaging of the metallic surface.[2] The heterogeneity of the refractive index of the metallic surface imparts high contrast images, caused by the shift in the resonance angle. [1] SPRM can achieve a thickness sensitivity of few tenths of nanometer and lateral resolution achieves values of micrometer scale. [3] SPRM is used to characterize surfaces such as self-assembled monolayers, multilayer films, metal nanoparticles, oligonucleotide arrays, and binding and reduction reactions. [4] [5] [6] [7] [8] Surface plasmon polaritons are surface electromagnetic waves coupled to oscillating free electrons of a metallic surface that propagate along a metal/dielectric interface. [9] Since polaritons are highly sensitive to small changes in the refractive index of the metallic material, [10] it can be used as a biosensing tool that does not require labeling. SPRM measurements can be made in real-time. [11] Wang and collaborators studied the binding kinetics of membrane proteins in single cells. [12]

History

The concept of classical SPR has been since 1968 but the SPR imaging technique was introduced in 1988 by Rothenhausler and Knoll. Capturing a high resolution image of low contrast samples for optical measuring techniques is a near impossible task until the introduction of SPRM technique that came into existence in the year 1988. In SPRM technique, plasmon surface polariton (PSP) waves are used for illumination. [13] In simple words, SPRI technology is an advanced version of classical SPR analysis, where the sample is monitored without label through the use of a CCD camera. The SPRI technology with the aid of CCD camera gives advantage of recording the sensograms and SPR images, and simultaneously analyzes hundreds of interactions. [14]

Basic principles

Surface plasmons or surface plasmon polaritons are generated by coupling of electrical field with free electrons in a metal. [15] [16] SPR waves propagate along the interface between dielectrics and a conducting layer rich in free electrons. [17]

As shown in Figure 2, when light passes from a medium of high refractive index to a second medium with a lower refractive index, the light is totally reflected under certain conditions. [18]

In order to get Total Internal Reflection (TIR), the θ1 and θ2 should be within a certain range that can be explained through the Snell’s law. When light passes through a high refractive index media to a lower refractive media, it is reflected at an angle θ2, which is defined in Equation 1.

θ2 = sin−1 ((η12)sinθ1) Eq. 1

 SPRM Image 2.
Figure 2. Total Internal Reflection (TIR), an incident beam of light with angle θ1, is passing through a media with refractive index η1, the beam reflects back at angle θ2, when the refractive index of the medium changes to a value η2.

In the TIR process some portion of the reflected light leaks a small portion of electrical field intensity into medium 2 (η1 > η2). The light leaked into the medium 2 penetrates as an evanescent wave. The intensity and penetration depth of the evanescent wave can be calculated according to Equations 2 and 3, respectively. [19]

Iz = I0 exp-z⁄dp Eq. 2

dp = λ/(4πη12 sin2θ122) Eq. 3

Figure 3 shows a schematic representation of surface plasmons coupled to electron density oscillations. The light wave is trapped on the surface of the metal layer by collective coupling to the electrons of the metal surface. When the electron’s plasma and the electric field of the wave light couple their frequency oscillations they enters into resonance. [20] [21]

 SPRM Image 3.
Figure 3. Cartoon of polaritons propagation along a metal dielectric interface, rich and poor electron density regions are referred as + and –, respectively.

Recently, the leakage light inside of the metal surface had been imaged. [22] Radiation of different wavelengths (green, red and blue) was converted into surface plasmon polaritons, through the interaction of the photons at the metal/dielectric interface. Two different metal surfaces were used; gold and silver. The propagation length of the SPP along the x-y plane (metal plane) in each metal and photon wavelength were compared. The propagation length is defined as the distance traveled by the SPP along the metal before its intensity decreases by a factor of 1/e, as defined in Equation 4.


Figure 4 shows the leakage light captured by a color CCD camera, of the green, red and blue photons in gold (a) and silver (b) films.In part c) of Figure 4, the intensity of the surface plasmon polaritons with the distance is shown. It was determined that the leakage light intensity is proportional to the intensity in the waveguide.

δSPP ≈ λ0 {(εm')2/(2πεm")} Eq. 4

Where δSPP is the propagation length; ε’m and ε’’m are the relative permittivity of the metal and λ0 is the free space wavelength. [23]
The metallic film is capable of absorbing light due to the coherent oscillation of the conduction band electrons induced by the interaction with an electromagnetic field. [24] Electrons in the conduction band induce polarization after interaction with the electric field of the radiation. A net charge difference is created in the surface of the metal film, creating a collective dipolar oscillation of electrons with the same phase. [25] When the electron motion matches the frequency of the electromagnetic field, the absorption of incident radiation occurs. The oscillation frequency of gold surface plasmons is found in the visible region of the electromagnetic spectrum, giving a red color while silver gives yellow color. [26] Nanorods exhibit two absorption peaks in the UV-vis region due to longitudinal and transversal oscillation, for gold nanorods the transverse oscillation generates a peak at 520 nm, while the longitudinal oscillation generates absorption at longer wavelengths, within a range of 600 to 800 nm. [26] [27] Silver nanoparticles shift their light absorption wavelengths to higher energy levels, where the blue shifting goes from 408 nm to 380 nm, and 372 nm, when they change from sphere to rod and wire, respectively. [28] The absorption intensity and wavelength of gold and silver depends on the size and shape of the particles. [29]
In Figure 5, the size and shape of silver nanoparticles influenced the intensity of the scattered light and maximum wavelength of silver nanoparticles. The triangular shaped particles appear red with a maximum scattered light at 670-680 nm, the pentagonal particles appear in green (620-630 nm) and the spherical particles have higher absorption energies (440-450 nm), appear in blue. [30]


Plasmon excitation methods

Surface plasmon polaritons are quasiparticles, composed by electromagnetic waves-coupled to free electrons of the conduction band of metals. [31] One of widely used methods uses to couple p-polarized light with the metal-dielectric interface is prism based coupling. [32] Prism couplers are the most widely used to excite surface plasmon polaritons. This method is also called Kretschmann-Raether configuration, where TIR creates an evanescent wave that couples the free electrons of the metal surface. [33] Waveguide coupling is also used to create surface plasmons.

Prism Coupling

Kretschmann-Raether configuration is used to achieve resonance between light and free electrons of the metal surface. In this configuration a prism with high refractive index is interfaced with a metal film. Light from a source propagates through the prism is made incident on the metal film. As a consequence of the TIR, some leaked through metal film, forming evanescent wave in the dielectric medium as in Figure 6. [11] The evanescent wave penetrates a characteristic distance into the less optical dense medium where it is attenuated. [34]


Figure 6 shows the Kretschmann-Raether configuration, where a prism with refractive index of η1 is coupled to a dielectric surface with a refractive index η2, the incidence angle of the light is θ.

The interaction between the light and the surface polaritons in the TIR can be explained by using the Fresnel multilayer reflection; the amplitude reflection coefficient (r_pmd) is expressed as follows in Equation 5. [35]

rpmd = {[rpm + rmd e(2ikmxq)]/[1+ rpm rmd e(2ik_mx q)]} Eq. 5

The power reflection coefficient R is defined as follows:

R=‖rpmd2 Eq. 6

In Figure 7, a schematic representation of the Otto prism coupling prism is shown. In the Figure 7, the air gap was shown a little thick just to explain the schematic although in reality, the air gap is so thin between prism and metal layer.


Waveguide coupling

The electromagnetic waves are conducted through an optical waveguide. When light enters to the region with a thin metal layer, it evanescently penetrates through the metal layer exciting a Surface Plasmon Wave (SPW). In waveguide coupling configuration, the waveguide is created when the refraction index of the grating is greater than that of substrate. Incident radiation propagates along the waveguide layer with high refractive index. [36] In Figure 8, electromagnetic waves are guided through a wave-guiding layer, once the optical waves reached the interface wave-guiding layer- metal an evanescent wave is created. The evanescent wave excites the Surface Plasmon at the metal-dielectric interface. [37]


Grating coupling

Due to the periodic grating, the phase matching between the incident light and the guide mode is easy to obtain. [38] According to Equation 7, the propagation vector (Kz) in the z direction can be tuned by changing the periodicity Λ. The grating vector can be modified, and the angle of resonant excitation can be controlled. [39] In Figure 9, q is the diffraction order it can have values of any integer (positive, negative or zero). [40]

Kz = q {2π/Λ} Eq.7


Resonance measurement methods

The propagation constant of a monochromatic beam of light parallel to the surface is defined by Equation 8. [41]
kx = [2π/λ] np sin θ Eq. 8
Where θ is the resonance angle for total internal reflection, k_sp is the propagation constant of the surface plasmon, and n_(p ) is the refractive index of the prism. When the wave vector of the SPW, ksp matches the wave vector of the incident light kx, SPW is expressed as: [41]

ksp = [2π/λ] md)/(εmd) Eq. 9
εd and εm represent the dielectric constant of dielectrics and the metal while the wavelength of the incident light corresponds to λ. kx and ksp can be represented as: [41]

θsp= sin−1  mεd/(εpm + εd)] Eq. 10

The surface plasmons are evanescent waves that have their maximum intensity at the interface and decay exponentially away from the phase boundary to a penetration depth. [12] The propagation of the surface plasmons is intensely affected by a thin film coating on the conducting layer. The resonance angle (θ) shifts, when the metal surface is coated with a dielectric material, due to the change of the propagation vector (k) of the surface plasmon. [42] This sensitivity is due to the shallow penetration depth of the evanescent wave. Materials with a high amount of free electrons are used. Metal films of roughly 50 nm made of copper, titanium, chromium and gold are used. However, Au is the most common metal used in SPR as well as in SPRM.
Scanning angle SPR is the most widely used method for detecting biomolecular interactions. [37] It measures the reflectance percentage (%R) from a prism/metal film assembly as a function of the incident angle at a fixed excitation wavelength. When the angle of incidence matches the propagation constant of the interface, this mode is excited at expenditure of the reflected light. As a consequence the reflectivity value at the resonance angle is dumped. [43]

The propagation constant of the polaritons can be modified by varying the dielectric material. This modification causes resonance angle shifting as in the example shown in Figure 10, from θ1 to θ2 due to the change on the surface plasmon propagation constant.


The resonance angle can be found by using Equation 11.
θSPR = sin−1 [(1/n1)(n22 ng2)/(n22 + ng2 )] Eq. 11

Where n_1 is n_2 and n_g are the refractive index of medium 1, 2 and the metal layer, respectively. [43]
Using TIR two-dimensional imaging is possible to achieve spatial differences in %R at a fixed angle (θ). A beam of monochromatic light is used to irradiate the sample at a fixed incident angle. The SPR image is created from the reflected light detected by a CCD camera. [12] The minimum value of % R at the resonance angle provides SPRM. [7]
Huang and collaborators developed a microscope with an objective with high numerical aperture (NA), which improve the lateral resolution at expense of the longitudinal resolution. [44]


Lateral resolution

The resolution of a conventional light microscopy is limited by the light diffraction limit. In SPRM, the excited surface plasmons adopt a horizontal configuration from the incident beam light. The polaritons will travel along the metal-dielectric interface, for a determined period, until they decay back into photons. Therefore, the resolution achieve by SPRM is determined by the propagation length ksp of the surface plasmons parallel to the incident plane. [43] The separation between to areas should be approximately the magnitude of ksp in order to be resolved. Berger, Kooyman and Greve showed that the lateral resolution can be tuned by changing the excitation wavelength, the better resolution is achieved when the excitation energy increases. Equations 4 and 12 defines the magnitude of the wave vector of the surface plasmons. [45]

ksp = (2π/λ) [(n22 ng2)/(n22 + ng2)] Eq. 12

Where n2 the refractive index of medium 2 is, ng is the refractive index of the metal film and λ is the excitation wavelength. [43]

Instrumentation

The surface plasmon resonance microscopy is based on surface plasmon resonance and recording desired images of the structures present on the substrate using an instrument equipped with a CCD camera. In the past decade, SPR sensing has been demonstrated to be an exceedingly powerful technique and used quite extensively in the research and development of materials, biochemistry and pharmaceutical sciences. [46]

The SPRM instrument works with the combination of the following main components: source light (typically He-Ne laser), that further travels through a prism that is attached to a glass side, coated with a thin metal film (typically gold or silver), where the light beam reflects at the gold/solution interface at an angle greater than the critical angle. [1] The reflected light from the interface surface area is recorded by a CCD detector, and an image is recorded. Although the above-mentioned components are some important for SPRM, additional accessories such as polarizers, filters, beam expanders, focusing lenses, rotating stage, etc., similar to several imaging methods are installed and used in the instrumentation for an effective microscopic technique as demaded by the application. The Figure 12 shows a typical SPRM. Depending on the applications, and to optimize the imaging technique, the researchers modify this basic instrumentation with some design changes that even include altering the source beam. One of such design changes that resulted in a different SPRM is an objective-type as shown in the Figure 11 with some modification in the optical configuration. [44]


The SPRi systems are currently manufactured by well known biomedical instrumentation manufacturers such as, GE Life Sciences, HORIBA, Biosensing USA, etc. The cost of these SPRi’s from the above manufacturers start from USD 100k and reaches almost to USD 250k depending on the application, whilst a mini prototype can be even made with a least budget of USD 2k just to witness this technique on a smaller scale that was built by Yijun Tang et al. to demonstrate the same to the high school students. The Figure 13 below is the USD 2k budget SPRi system built by Yijun Tang et al. [47]


Sample preparation

To perform measurements for SPRM, the sample preparation is a critical step. There are two factors that can be affected by the immobilization step: one is the reliability and reproducibility of the acquire data. It is important to ensure stability to the recognition element; such as antibodies, proteins, enzymes, under the experiment conditions. Moreover, the stability of the immobilized specimens will affect the sensitivity, and/or the limit of detection (LOD). [48] [49]
One of the most popular immobilization methods used is Self-Assembled Monolayer (SAM) on gold surface. Jenkins and collaborators 2001, used mercaptoethanol patches surrounded by SAM composed of octadecanethiol (ODT) to study the adsorption of egg-phosphatidylcholine on the ODT SAM. [4] A pattern of ODT-mercaptoethanol was made onto a 50 nm gold film. The gold film was obtained through thermal evaporation on a LaSFN 9 glass. The lipid vesicles were deposited on the ODT SAM through adsorption, giving a final multilayer thickness greater than 80 Å.

11-Mercaptoundecanoic acid-Self assembled monolayer (MUA-SAM) were formed on Gold coated BK7 slides. A PDMS plate was masked on the MUA-SAM chip. Clenbuterol (CLEN) was attached to BSA molecules through amide bond, between the carboxylic group of BSA and the amine group of CLEN molecules. In order to immobilize BSA on the gold surface, the spots created through PDMS making were functionalized with sulfo-NHS and EDC, subsequently 1% BSA solution was poured in the spots and incubated for 1 hour. Non-immobilized BSA was rinsed out with PBS and CLEN solution was poured on the spots, unimmobilized CLEN was remove through PBS rinse. [50]

An alkanethiol-SAM was prepared in order to simultaneously measure the concentration of horseradish peroxidase (Px), Human Immunoglobulin E (IgE), Human choriogonadotropin (hCG) and Human immunoglobulin G (IgG), through SPR. The alkanethiols made of carbon chains composed by 11 and 16 carbons were self-assembled on the sensor chip. The antibodies were attached to the C16 alakanethiol, which had a terminal carboxylic group. [51]


The micro patterned electrode was fabricated by gold deposition on microscope slides. PDMS stamping was used to produce an array of hydrophilic/hydrophobic surface; ODT treatment followed by immersion in 2-mercaptoethanol solutions rendered a functionalized surface for lipid membranes deposition. The patterned electrode was characterized through SPRM. In the Figure 14 B, the SPRM image reveals the size of the pockets, which was 100 um x 100 um, and they were 200 um apart. As is seen in the image the remarkable contrast of the image is due to the high sensitivity of the technique.

Applications

SPRM is a useful technique for measuring concentration of biomolecules in the solution, detection of binding molecules and real time monitoring of molecular interactions. It can be used as biosensor for surface interactions of biological molecules: antigen-antibody binding, mapping and sorption kinetics. For example, one of the possible reason of Type 1 diabetes of children is the high level presence of Cow’s milk antibodies IgG, IgA, IgM (mainly due to IgA) in their serum. [52] Cow’s milk antibodies can be detected in the milk and serum sample using SPRM. [53] SPRM is also advantageous to detect the site-specific attachment of lymphocyte B or T on antibody array. This technique is convenient to study the label free and real time interactions of cells on the surface. So SPRM can be served as diagnostic tool for cell surface adhesion kinetics. [54] Besides its merits, there are limitations of SPRM though. It’s not applicable for detecting low molecular weight molecules. Although it’s label free but will need to have crystal clean experimental conditions. Sensitivity of SPRM can be improved with coupling of MALDI-MS. [55] There are number of applications of SPRM from which some of them are being described here.

Membrane proteins

Membrane proteins are responsible for the regulation of cellular responses to extracellular signals. It has been the challenging thing to investigate the involvement of membrane proteins in disease biomarkers and therapeutic targets and its binding kinetics with their ligands. Traditional approaches could not reflect clear structures and functions of membrane proteins. [12] In order to understand the structural details of membrane proteins, there is a need of alternate analytical tool, which can provide three-dimensional and sequential resolutions that can monitor membrane proteins. Atomic force microscopy (AFM) is an excellent method for obtaining high spatial resolution images of membrane proteins, [56] but it might not be helpful to investigate its binding kinetics. Fluorescence-based microscopy (FLM) can be used to study the interactions of membrane proteins in individual cells but it requires development of proper labels and needs tactics for different target proteins. [57] Furthermore, host protein may be affected by the labeling. [58]

Binding kinetics of MP’s in the single living cells can be studied via label free imaging method based on SPR Microscopy without extracting the proteins from the cell membranes, which help scientists to work with the actual conformations of the membrane proteins. Furthermore, distribution and local binding activities of membrane proteins in each cell can be mapped and calculated. SPR microscopy (SPRM) makes possible to simultaneously optical and fluorescence imaging of the same sample, which prove to get the advantages of both label-based and label-free detection methods in the single setup. [44] [59]

Detection of DNA hybridization

SPR imaging is used to study the multiple adsorption interactions in an array format under same experimental conditions. Nelson and his coworkers introduced a multistep procedure to create DNA arrays on gold surfaces for use with SPR imaging. [60] Affinity interactions can be studied for a variety of target molecules e.g. proteins and nucleic acids. Mismatching of bases in the DNA sequence leads to the number of lethal diseases like lynch syndrome which has high risk of colon cancer. [61]

SPR imaging is useful to monitor adsorption of molecules on the gold surface which is possible because of the change in the reflectivity from the surface. First G-G mismatch pair is stabilized by attaching it with the ligand, naphthyridine dimer, through hydrogen bonding which make the hairpin structures in double stranded DNA on gold surface. Binding of Dimer with DNA enhances the free energy of hybridization, which causes change in index of refraction. [62]

SPRM Image 15.
Figure 15. Structure of the G–G mismatch stabilizing the naphthyridine dimer (blue) is shown hydrogen bonding to two guanine bases (black). [62]

DNA array is fabricated to test the G–G mismatch stabilizing properties of the naphthyridine dimer. Each of the four immobilized sequences in the array differed by one base. The position of this base is indicated by an X in sequence 1 as shown in Figure 16. The SPR difference image is only detected for the sequence having cytosine (C) base at the X position in sequence 1, the complementary sequence to sequence 2. However, the SPR difference image corresponding to the addition of sequence 2 in the presence of the naphthyridine dimer shows that, in addition to its complement, sequence 2 also hybridizes to the sequence that forms a G–G mismatch. These results demonstrate that SPR imaging is a promising tool for monitoring single base mismatches and screen out the hybridized molecules. [62]


Antibody binding to protein arrays

Vishal Kanda et al. demonstrates SPR imaging to study the binding of antibodies to protein array. [63] Amine functionalities on the gold surface with proteins array, is used to study binding of antibodies. Immobilization of the protein was done by flowing protein solutions through the PDMS micro channels. Then PDMS was removed from the surface and solutions of antibody were flowed over the array. Three-component protein array containing the proteins human fibrinogen, ovalbumin, and bovine IgG is shown in Figure 17, SPR images obtained by Kariuki and co-workers. This contrast in the array is due to difference of refractive index which is outcome of local binding of antibodies. These images show that there is a high degree of antibody binding specificity and a small degree of non-specific adsorption of the antibody to the array background, which can be improved to modify the array background. Based on these results, SPR imaging technique can be opted as diagnostic tool for studying the antibody interactions to protein arrays. [63] [64]


SPRM Coupled With MS

Discovery and validation of protein biomarkers are crucial for diseases diagnosis. Coupling of SPRM with MALDI-mass spectrometer (SUPRA-MS) enables the multiplex quantification of binding and molecular characterization on the basis of different masses. SUPRA-MS is used to detect, identify and characterize the potential breast cancer biomarker, LAG3 protein, introduced in the human plasma. Glass slides were taken to prepare gold chips via coating with thin layers of chromium and gold by sputtering process. Gold surface was functionalized using solution of 11-Mercapto-1-undecanol (11-MUOH) and 16-mercapto-1-hexadecanoic acid (16-MHA). This self-assembled monolayer was activated with sulfo-NHS and EDC. Pattern of sixteen droplets was deposited on the macroarray. Immunoglobin G antibodies were spotted against Lymphocyte activation gene 3 (α-LAG3) and rat serum albumin (α-RSA). After placing biochip in the SPRi and running buffer solution in the flow cell, α-LAG3 was injected. Special image station was used on the proteins that are attached. This station can also be placed on the MALDI. Before placing on the MALDI, captured proteins were reduced, digested and loaded with matrix in order to avoid contamination. [55]


Antigen density is directly proportional to change in reflectivity ΔR because evanescent wave penetration depth Lzc is larger than thickness of immobilized antigen layer.[65]

Antigen Density = ᴦ (pg/mm2) = {(∆R×Lzc)/[(SPR×(∂n⁄∂c)]} Eq. 13

∂n/∂C is the index increment of the molecule and SPR is the sensitivity prism, reflectivity.

Clean mass spectrum was obtained for LAG3 protein due to good tryptic digestion and homogeneity of the matrix (α-cyano-4-hydroxycinnamic acid). Relatively high intensity m/z peak of LAG3 protein was found at 1,422.70amu with average mascot score of 87.9 ± 2.4. Validation of MS results was further confirmed by MS-MS analysis. These results are similar to classical analytical method in-gel digestion. [55]

Greater S/N>10, 100% reliability and detection at femtomole level on chip proves the credibility of this coupling technique. One can find protein-protein interaction and on-chip peptide distribution with high spatial resolution using subjected technique.[55]

DNA aptamers

Aptamers are particular DNA ligands that target biomolecules such as proteins. SPR imaging platform would be a good choice to characterize aptamer -protein interactions. To study the aptamer-protein interaction, first oligonucleotides are grafted through formation of thiol Self Assembling Monolayer (SAM) on gold substrate using piezoelectric dispensing system. Thiol groups are introduced on DNA nucleotides by N-hydroxysuccinimide (NHS). Target oligonucleotides having a primary amine group at their 59th end are conjugated to HS-C (11)-NHS in phosphate buffer solution at pH 8.0 for one hour at room temperature. Aptamer grafting biosensor is placed on SPRM after rinsing. Then Thrombin is co-injected with excess of cytochrome C for signal specificity. Concentration of free thrombin is determined by calibration curve obtained by plotting initial slope of the signal at the beginning of injection against concentration. The interaction of thrombin and the aptamer can be monitored on microarray in real-time during injections of thrombin at different concentrations. Solution phase dissociation constant KDsol (3.16±1.16 nM) is calculated from the measured concentrations of free thrombin.

KDsol = [THR][APT]⁄[THR⋯APT] Eq.14

[THR---APT] = cTHR – [THR], the equilibrium concentration of thrombin attached to aptamers in solution and [APT] = cAPT – [THR---APT], the concentration of free aptamers in solution.

Surface phase dissociation constant KDsurf (3.84±0.68) is obtained by fitting Langmuir adsorption isotherm on equilibrium signals. Both dissociation constants are significantly different because KDsurf is dependent on the surface grafting density as shown in Figure 19. This dependence extrapolates linearly at low sigma to solution-phase affinity.

The difference in SPRi image can gives us information regarding the presence of binding and specificity but not suitable for quantification of free protein in case of multiple affinity sites. The real time monitoring of the interaction is possible by using SPRM to study the kinetics and the affinity of the interactions. [66]


Detection of polymer interaction

Despite of using surface plasmon resonance imaging (SPRi) in biology to characterize interactions between two biological molecules, it is also useful to monitor the interactions between two polymers. In this approach, one polymer, called as host protein HP, is immobilized on the surface of a biochip and the other polymer designated as guest polymer GP is inserted on the SPRi-Biochip to study the interactions. Vollmer et al., used amine functionalized poly (β-cyclodextrin), which acts as Host protein. This poly(β-CD) NH2 can be synthesized from condensation reaction between β-cyclodextrins and epichlorohydrin. [67] Guest protein, PEG (ada)4, was poly (ethylene glycol) (PEG) having four arms with adamantyl (ada) functionality at the end. [68] SPRi biochip was used for immobilization of HP of different concentrations. An array of HP active sites was produced on the chip. The attachment of HP was done through its amino groups to N-hydroxy succinimide functionalities on the gold surface. First SPRi system was filled running buffer solution followed by placing of SPRi –biochip into the analysis chamber. Two solutions of different concentrations of GP was 1g/L and0.1 g/L were injected in the flow cell. The association and the dissociation of both polymers can be monitored in real-time on the basis of change in reflectivity and images from SPRM can be differentiated on the basis of white spots (association phase) and black spots (dissociation phase). PEG without adamantyl groups didn’t show adsorption on β-cyclodextrin cavities. On the other hand, there wasn’t any adsorption of GP without HP on the chip. Change in SPRi response on the reaction sites is provided by the capturing of kinetic curves and real time images from the CCD camera. Local changes in light reflectivity are directly related to quantity of target molecules on each point. Variation at the surface of the chip provide comprehensive knowledge on molecular binding and kinetic processes. [69]

Role of SPRM in bio-mineralization

One of the important class of biomaterials is polymer hydroxyapatite that is remarkably useful in the field of bone regeneration because of its resemblance with natural bone material. The advantage of hydroxyapatite, (Ca10(PO4)6(OH)2, is being started to form inside the bone tissue through mineralization which also advocate the enhancement of osteointegration. Biomineralization is also called calcification, in which calcium cations come from cells and physiological fluids while phosphate anions are produced from hydrolysis of phosphoesters and phosphoproteins as well as from the body fluids. This phenomenon is also tested in vitro studies. [70]

For vitro studies, Polyamidoamine (PAMAM) dendrimers with amino- and carboxylic acid- external reactive shells are considered as sensing phase. These dendrimers are required to immobilized on the gold surface and inactive to gold surface. Hence, thiols groups have to be introduced at the terminals of dendrimers so that dendrimers can be attached on the gold surface. Carboxylic groups are functionalized by N,N-(3-dimethylaminopropyl)-N’-ethyl-carbodiimide hydrochloride (EDC) and N-hydroxysuccinimide (NHS) solutions in phosphate buffer. Functional groups (amide, amino and carboxyl) act as ionic pumps capturing calcium ions from the test fluids; then calcium cations binds with phosphate anions to generate calcium-phosphate mineral nuclei on the dendrimer surface. [70]

SPRM is expected to be sensitive enough to provide important quantitative information on mineralization’s occurrence and kinetics. This detection of the mineralization is based on the specific mass change induced by the mineral nuclei formation and growth. Nucleation and progress in mineralization can be monitored by SPRM as shown in Figure 20. PAMAM-containing sensors are fixed on the SPRi analysis platform and then exposed to experimental fluids in the flow cell as shown in the Figure 21. SPRM is not adapted to sense the origin and nature of mass change but it detects the modification of refractive index due to mineral precipitation. [70]


Conclusion

SPRM is very promising optical label free technique in which polarized light is incident on prism covered by thin gold layer at its rear side. Free electrons at the surface of the chip show surface plasmon resonance by absorbing incident photons and create plasmon waves depending upon particular incident angle and wavelength. Probe molecules are immobilized on the chip, and target molecules are allowed to interact with probe molecules which causes the change in reflectivity captured by CCD camera. This technique is being used for monitoring the kinetic behaviour, surface interaction of biomolecules, polymers, mismatch based pairs, electrochemical and bio mineral detection. It is coupled with MALDI-MS to make it more useful. Being simple and versatile, this technique may lead to more physical and biomedical applications.

References

  1. 1 2 3 Campbell, C, T; Kim, G (2007). "SPR microscopy and its applications to high throughput analyses of biomolecular binding events and their kinetics". Biomaterials 28: 2380–2392. doi:10.1016/j.biomaterials.2007.01.047.
  2. Peterson, A, W; Halter, M; Tona, A; Plant, A, L (2014). "High resolution surface plasmon resonance imaging of single cells". BMC Cell Biology 15 (35): 2–14. doi:10.1186/1471-2121-15-35.
  3. Hickel, W; Kamp, D; Knoll, W (1989). "Surface Plasmon Microscopy". Nature 339: 186. doi:10.1038/339186a0.
  4. 1 2 Jenkins, A; Neumann, T; Offenhausser, A (2001). "Surface Plasmon Microscopy Measurements of Lipid Vesicle Adsorption ona Micropatterned Self-Assembled Monolayer". Langmuir 17: 265–267. doi:10.1021/la991680q.
  5. Nicoletti, O; De La Pena, F; Leary, R, K; Holland, D, J; Ducati, C; Midgley, P, A (2013). "Three-dimensional imaging of localized surface plasmon resonances of metal nanoparticles". Nature 502: 80–84. doi:10.1038/nature12469.
  6. Thiel, A, J; Frutos, A, G; Jordan, C, E; Corn, R, M; Smith, L, M (1997). "In situ Surface Plasmon Resonance Imaging Detection on DNA Hybridization to Oligonucleotide Arrays on Gold Surfaces". Analytical Chemistry 69: 4948–4956. doi:10.1021/ac9708001.
  7. 1 2 Zizisperger, M; Knoll, W (1998). "Multispot parallel on-line monitoring of interfacial binding reactions by surface plasmon microscopy". Progress in Colloid & Polymer Science 109: 244–253. doi:10.1007/bfb0118177.
  8. Flatgen, G; Krischer, K; Ertl, G (1996). "Spatio-temporal pattern formation during the reduction of peroxodisulfate in the bistable and oscillatory regime: a surface plasmon microscopy study". Journal of Electroanalytical Chemistry 409: 183–194. doi:10.1016/0022-0728(96)04511-1.
  9. Agranovich, V, M; Mills, D, L (1982). Surface Polaritons - Electromagnetic Waves at Surfaces and Interfaces. New York, N.Y: North-Holland.
  10. Yang, X, Y; Xie, W, C; Liu, D, M (2008). "Design of Highly Sensitive Surface Plasmon Resonance Sensors Using Planar MEtallic Films Closely Coupled to Nanogratings". Chinese Physical Letters 25 (1): 148–151. doi:10.1088/0256-307x/25/1/041.
  11. 1 2 Tang, Y; Zeng, X; Liang, J (2010). "Surface Plasmon Resonance: An Introduction of a Surface Spectroscopy Technique". Journal of Chemical Education 87 (7): 742–746. doi:10.1021/ed100186y.
  12. 1 2 3 4 Wei, W; Yunze, Y; Shaopeng, W; Vinay, J, N; Qiang, L; Jie, W; Nongjian, T (2012). "Label-free measuring and mapping of binding kinetics of membrane proteins in single living cells". Nature Chemistry 4: 846–853. doi:10.1038/nchem.1434.
  13. Rothenhausler, B; Knoll, W (1988). "Surface plasmon microscopy". Nature 332: 615–617. doi:10.1038/332615a0.
  14. Spoto, G; Minunni, M (2012). "Surface plasmon resonance imaging: What Next?". The Journal of Physical Chemistry Letters 3: 2682–2691. doi:10.1021/jz301053n.
  15. Mills, D, L; Burstein, E (1974). "Polaritons: the electromagnetic modes of media". Reports on Progress in Physics 37 (7): 817–926. doi:10.1088/0034-4885/37/7/001.
  16. Klotzbach, U; Lasagni, A, F; Panzner, M; Volker, F (2011). "Fabrication and Characterization in the Micro-Nano Range". Advanced Structure Materials 10: 29–46.
  17. Plasmonic Nanoguides and circuits: Nanophotonic Components Utilizing Channel Plasmon Polaritons. Singapore: Pan Stanford. 2009.
  18. Ho, H, P; Wu, S, Y (2007). Nanotechnology in Biology and Medicine MEthods, Devices and Applications: Biomolecule Sensing Using Surface Plasmon Resonance. Boca Raton: Taylor & Francis Group. p. 792.
  19. Toomre, D; Manstein, D, J (2001). "Lighting up the cell surface with evanescent wave microscopy". TRENDS in cell Biology 11 (7): 298–303. doi:10.1016/s0962-8924(01)02027-x.
  20. Barnes, W, L; dereux, A; Ebbesen, T, W (2003). "Surface plasmon subwavelength optics". Nature 424: 824–830. doi:10.1038/nature01937. PMID 12917696.
  21. Benson, O (2011). "Assembly of hybrid photonic architectures from nanophotonic constituents". Nature 480: 193–199. doi:10.1038/nature10610.
  22. Pan, M, Y; Lin, E, H; Wang, L; Wei, P, K (2014). "Spectral and mode properties of surface plasmon polariton waveguides studied by near-field excitation and leakage-mode radiation measurement". Nanoscale Research Letters 9 (1): 430. doi:10.1186/1556-276x-9-430.
  23. Barnes, W, L (2006). "Surface Plasmon-polariton length scales: a route to sub-wavelength optics". Journal of Optics A: Pure and Applied Optics 8: 87–93. doi:10.1088/1464-4258/8/4/S06.
  24. Mulvaney, P (1996). "Surface Plasmon Spectroscopy of Nanosized Metal Particles". Langmuir 12: 788–800. doi:10.1021/la9502711.
  25. Tiwari, A; Narayan, J (2005). Nanoingeneering of Structural, Functional and Smart Materials: Self-Assembled Au Nanodots in a ZnO Matrix: A Novel Way to Enhance Electrical and Optical Characteristics of ZnO Films. Boca Raton: Taylor & Francis Group. p. 740.
  26. 1 2 Eustis, S; El Sayed, M (2006). "Why gold nanoparticles are more precious then pretty gold: Noble metal surface plasmon resonance and its enhancement of the radiative and nonradiative properties of nanocrystals of different shapes". Chemical Society Reviews 35: 209–217. doi:10.1039/b514191e.
  27. Kim, F, H; Song, J; Yang, P (2002). "Photochemical Synthesis of Gold Nanorods". Journal of the American Chemical Society 124: 14316–14317. doi:10.1021/ja028110o.
  28. Hu, J, G; Chen, Q; Xie, Z, X; Han, G, B; Wang, R, H; Ren, B; Zhang, Y; Yang, Z, L; Tian, Z, Q (2004). "A Simple and Effective Route for the Synthesis of Crystalline Silver Nanorods and Nanowires". Advanced Functional Materials 14 (2): 183–189. doi:10.1002/adfm.200304421.
  29. Anker, J, N; Hall, W, P; Lyandres, O; Shah, N, C; Zhao, J; Van Duyne, R, P (2008). "Biosensing with plasmonic nanosensors". Nature Materials 7: 442–453. doi:10.1038/nmat2162. PMID 18497851.
  30. Mock, J, J; Barbic, M; Smith, D, R; Schultz, D, A; Schultz, S (2002). "Shape effects in plasmon resonance of individual colloidal silver nanoparticles". The Journal of Chemical Physics 116: 6755–6759. doi:10.1063/1.1462610.
  31. Karpinski, P; Miniewicz, A (2011). "Surface Plasmon Polariton Excitation in Metallic Layer Via Surface Relief Gratings in Photoactive Polymer Studied by the Finite-Difference Time Domain Method". Plasmonics 6: 541–546. doi:10.1007/s11468-011-9234-3.
  32. Mukherji, S; Hossain, M, I; Kundu, T; Chandratre, D (2014). Micro and Smart Deices an Systems: Development of a Surface Plasmon Resonance-Based Biosensing System. Springer.
  33. Kreyschmann, E; Raether, H (1968). "Radiative Decay of Non Radiative Surface Plasmons Excited by Light". Z. Naturforsch. 23a: 2135–2136. doi:10.1515/zna-1968-1247.
  34. Sekhar, P, K; Uwizeye, V (2012). MEMS for biomedical applications: Review of sensor and actuator mechanisms for bioMEMS. Cambridge: Cambridge.
  35. Homola, J (2006). Surface Plasmons Based Sensors: Electromagnetic Theory of Surface Plasmons. Springer. pp. 3–44.
  36. Schimitt, K; Hoffmann, C (2010). High - Refractive - Index waveguide Platforms for Chemical and Biosensing. Springer-Verlag.
  37. 1 2 Homola, J (2003). "Present and future of surface plasmon resonance biosensors". Analytical and Bioanalytical Chemistry 377: 528–539. doi:10.1007/s00216-003-2101-0.
  38. Solgaard, O (2009). Photonic Microsystems Micro and Nanotechnology Applied to Optical Devices and Systems. Springer.
  39. Knoll, W; Kasry, A; Liu, J; Neumann, T; Niu, L; Park, H; Paulsen, H; Robelek, R; Yao, D; Yu, F (2008). Handbook of Surface Plasmon Resonance: Surface Plasmon Fluorescence Techniques for Bioaffinity Studies. RSC Publishing.
  40. Baba, A; Kaneko, F; Advincula, R; Knoll, W (2011). Functional Polymer Films: 2 Volume Set: Electrochemical Surface Plasmon Resonance Methods for Polymer Thin Films. weinheim: Wiley -VCH Verlag GmbH & Co. p. 1128.
  41. 1 2 3 Gwon, H, R; Lee, S, H (2010). "Spectral and angular responses of surface plasmon resonance based on the Kretschmann prism configuration". Materials Transactions 51 (6): 1150–1155. doi:10.2320/matertrans.m2010003.
  42. Gupta, B, D; Jha, R (2015). Handbook of Optical Sensors: Surface Plasmon Measurement: Principles and Techniques. Boca Raton: Taylor & Francis Group.
  43. 1 2 3 4 Giebel, K, F; Bechinger, C; Herminghaus, S; Riedel, M; Leiderer, P; Weiland, U; Bastmeyer, M (1999). "Imaging of Cell-Substrate Contacts of Living Cells with Surface Plasmon Microscopy". Biophysical Journal 76: 509–516. doi:10.1016/s0006-3495(99)77219-x.
  44. 1 2 3 Huang, B; Yu, F; Zare, R, N (2007). "Surface plasmon resonance imaging using a high numerical aperture microscope objective". Analytical Chemistry 79: 2979–2983. doi:10.1021/ac062284x.
  45. Berger, C, E, H; Kooyman, R, P, H; Greve, J (1994). "Resolution in Surface Plasmon Microscopy". REVIEW OF SCIENTIFIC INSTRUMENTS 65 (9): 2829–2836. doi:10.1063/1.1144623.
  46. Shumaker-Parry, J; Campbell, C, T (2004). "Quantitative Methods for Spatially-Resolved Adsorption/Desorption Measurements in Real Time by SPR Microscopy". Analytical Chemistry 76 (4): 907–917. doi:10.1021/ac034962a.
  47. Yijun, T; Xiangqun, Z; Jennifer, L (2010). "Surface Plasmon Resonance: An Introduction to a Surface Spectroscopy Technique". Journal of Chemical Education 87 (7): 742–746. doi:10.1021/ed100186y.
  48. Scarano, S; Mascini, M; Turner, A; Minunni, M (2010). "Surface plasmon resonance imaging for affinity based biosensors". Biosensors Bioelectronics 25 (5): 957–966. doi:10.1016/j.bios.2009.08.039.
  49. Homola, J (2008). "Surface plasmon resonance sensors for detection of chemical and biological species". Chemical Reviews 108: 462–493. doi:10.1021/cr068107d. PMID 18229953.
  50. Yao, M; Wu, Y; Fang, X; Yang, Y; Liu, H (2015). "Spectral surface plasmon resonance imaging for the detection of clenhuterol via three-dimensional immobilization bioprobes". Analytical Biochemistry 475: 40–43. doi:10.1016/j.ab.2015.01.012.
  51. Hamola, J; Vaisocherova, H; Dostalek, J; Pilarik, M (2005). "Multi-analyte surface plasmon resonance biosensing". Methods 37: 26–36. doi:10.1016/j.ymeth.2005.05.003.
  52. M. Virtanen S, T. Saukkonen, E. Savilahti, K. Ylönen, L. Räsänen, A. Aro, M. Knip, J. Tuomilehto and H. Akerblom, "Diet, cow's milk protein antibodies and the risk of IDDM in Finnish children. Childhood Diabetes in Finland Study Group," Diabetologia., pp. 37(4):381-7, 1994.
  53. Scarano, S; Scuffi, C; Mascini, M; Minunni, M (2011). "Surface Plasmon Resonance imaging-based sensing for anti-bovine immunoglobulins detection in human milk and serum.". Anal Chim Acta 707 (1-2): 178–183. doi:10.1016/j.aca.2011.09.012.
  54. Suraniti, E; Sollier, E; Calemczuk, R; Livache, T; Marche, P, N; Villersb, M, B; Roupioz, Y (2007). "Real-time detection of lymphocytes binding on an antibody chip using SPR imaging". Lab Chip 7: 1206–1208. doi:10.1039/b708292d.
  55. 1 2 3 4 Remy-Martin, F; El Osta, M; Luchhi, G; Zeggary, R; Leblois, T; Bellon; Ducoroy, P; Boireau, W (2012). "Surface plasmon resonance imaging in arrays coupled with mass spectrometry (SUPRA-MS): proof of concept of on-chip characterization of a potential breast cancer marher in human plasma". Analytical and Bioanalytical Chemistry 404 (2): 423–432. doi:10.1007/s00216-012-6130-4.
  56. Li, G, Y; Xi, N; Wang, D, H (2006). "Probing membrane proteins using atomic force micrsocopy". Journal of Cell Biochemistry 97: 1191–1197. doi:10.1002/jcb.20753.
  57. Groves, J, T; Parthasarathy, R; Forstner, M, B (2008). "Fluoroscence imaging of membrane dynamics". Annu. Rev. Biomed. Eng. 10: 311–338. doi:10.1146/annurev.bioeng.10.061807.160431.
  58. Johnson, A, E (2005). "Fluoroscence approaches for determining protein conformations, Iinteractions and mechanisms at membranes". Traffic 6: 1078–1092. doi:10.1111/j.1600-0854.2005.00340.x.
  59. Wang, W; Foley, K; Shan, X; Wang, S; Eaton, S; Nagraj, V, J; Wiktor, P; Patel, U; Tao, N (2011). "Single cells and intracellular processes studied by a plasmonic-based electrochemical impedance microscopy". Nature Chemistry 3: 249. doi:10.1038/nchem.961.
  60. Nelson, B, P; Grimsrud, T, E; Liles, M, R; Goodman, R, M; Corn, R, M (2001). "Surface Plasmon Resonance Imaging Measurements of DNA and RNA Hybridization Adsorption onto DNA Microarrays". Analytical Chemistry 73 (1): 1–7. doi:10.1021/ac0010431.
  61. Kastrinos, F; Mukherjee, B; Tayob, N; Wang, F; Sparr, J; Raymond, V, M; Bandipalliam, P; Stofell, E, M; Gruber, S, B; Syngal, S (2009). "Risk of pancreatic cancer in families with Lynch syndrome". JAMA 302 (16): 1790–1795. doi:10.1001/jama.2009.1529. PMID 19861671.
  62. 1 2 3 Smith, E, A; Kyo, M; Kumasawa, H; Nakatani, K; Saito, I; Corn, R, M (2002). "Chemically Induced Hairpin Formation in DNA Monolayers". Journal of the American Chemical Society 124: 6810–6811. doi:10.1021/ja026356n.
  63. 1 2 Kanda, V; Kariuki, J, K; Harrison, D, J; McDermott, M, T (2004). "Label-Free Reading of Microarray-Based Immunoassays with Surface Plasmon Resonance imaging". Analytical Chemistry 76 (24): 7257–7262. doi:10.1021/ac049318q.
  64. Kariuki, J, K; Kanda, V; Mc Dermott, M, T; Harrison, D, J (2002). Micro Total Analysis Systems. Nara: Kluwer Academic Publisher. pp. 230–232.
  65. E. Stenberg, B. Persson, H. Roos and Urbaniczky., J Colloids Interface Sci, p. 143:513–526, 1991.
  66. Daniel, C; Roupioz, Y; Gasparutti, D; Livache, T; Buhot, A (2013). "Solution-Phase vs Surface-Phase Aptamer-Protein Affinity from a Label-Free Kinetic Biosensor". PLOS ONE 8 (9): e75419. doi:10.1371/journal.pone.0075419.
  67. Renard, E; deratani, A; Volet, G; Sebille, B (1997). "Preparation and characterization of water soluble high molecular weight b-cyclodextrin-epichlorohydrin polymers". Eur. Polym. J. 33: 49–57. doi:10.1016/s0014-3057(96)00123-1.
  68. David, C; Millot, M, C; Brachais, L; Amiel, C; Sebille, B (2000). "Synthesis of hydrophobically end-capped poly(ethylene glycol)s with UV absorbing properties". Macromol. Rapid Commun. 21: 990–993. doi:10.1002/1521-3927(20000901)21:14<990::aid-marc990>3.3.co;2-2.
  69. Vollmer, N; Trombini, F; Hely, M; Bellon, S; Mercier, K; Cazeneuve, C (2015). "Methodology to study polymers interaction by surface plasmon resonance imaging". MethodsX 2: 14–18. doi:10.1016/j.mex.2014.12.001.
  70. 1 2 3 Izabela-Cristina, Stancua (2010). Intelligent and Biosensors: SPR Imaging Label-Free Control of Biomineral Nucleation!?, Intelligent and Biosensors. INTECH. ISBN 978-953-7619-58-9.