Ultra-high vacuum

Ultra-high vacuum is the vacuum regime characterised by pressures lower than about 10−7 pascal or 100 nanopascals (10−9 mbar, ~10−9 torr). UHV requires the use of unusual materials for equipment, and heating of the entire system above 100 °C for many hours ("baking") to remove water and other trace gases which adsorb on the surfaces of the chamber. At these low pressures the mean free path of a gas molecule is approximately 40 km, so gas molecules will collide with the chamber walls many times before colliding with each other. Almost all interactions therefore take place on various surfaces in the chamber.

Concepts involved

Material limitations

Materials which are not allowed due to high vapor pressure:

Technical limitations:

Typical uses for ultra-high vacuum

Ultra-high vacuum is necessary for many surface analytic techniques such as:

UHV is necessary for these applications to reduce surface contamination, by reducing the number of molecules reaching the sample over a given time period. At 0.1 mPa (10−6 Torr), it only takes 1 second to cover a surface with a contaminant, so much lower pressures are needed for long experiments.

UHV is also required for:

and, while not compulsory, can prove beneficial in applications such as:

Achieving ultra-high vacuum

Typically, UHV requires:

Outgassing is a problem for UHV systems. Outgassing can occur from two sources: surfaces and bulk materials. Outgassing from bulk materials is minimized by selection of materials with low vapor pressures (such as glass, stainless steel, and ceramics) for everything inside the system. Materials which are not generally considered absorbent can outgas, including most plastics and some metals. For example, vessels lined with a highly gas-permeable material such as palladium (which is a high-capacity hydrogen sponge) create special outgassing problems.

Outgassing from surfaces is a subtler problem. At extremely low pressures, more gas molecules are adsorbed on the walls than are floating in the chamber, so the total surface area inside a chamber is more important than its volume for reaching UHV. Water is a significant source of outgassing because a thin layer of water vapor rapidly adsorbs to everything whenever the chamber is opened to air. Water evaporates from surfaces too slowly to be fully removed at room temperature, but just fast enough to present a continuous level of background contamination. Removal of water and similar gases generally requires baking the UHV system at 200 to 400 °C while vacuum pumps are running. During chamber use, the walls of the chamber may be chilled using liquid nitrogen to reduce outgassing further.

Hydrogen and carbon monoxide are the most common background gases in a well-designed, well-baked UHV system. Both Hydrogen and CO diffuse out from the grain boundaries in stainless steel. Helium could diffuse through the steel and glass from the outside air, but the abundance of He is usually negligible in the atmosphere.

There is no single vacuum pump that can operate all the way from atmospheric pressure to ultra-high vacuum. Instead, a series of different pumps is used, according to the appropriate pressure range for each pump. Pumps commonly used to achieve UHV include:

UHV pressures are measured with an ion gauge, either a hot filament or an inverted magnetron type.

Metal seals, with knife edges on both sides cutting into a soft, copper gasket. This all-metal seal can maintain pressures down to 100 pPa (~10−12 Torr).

Measuring high vacuum

Main article: Pressure measurement

Measurement of high vacuum is done using a nonabsolute gauge that measures a pressure-related property of the vacuum, for example, its thermal conductivity. See, for example, Pacey.[1] These gauges must be calibrated.[2] The gauges capable of measuring the lowest pressures are magnetic gauges based upon the pressure dependence of the current in a spontaneous gas discharge in intersecting electric and magnetic fields.[3]

UHV manipulator

A UHV manipulator allows an object which is inside a vacuum chamber and under vacuum to be mechanically positioned. It may provide rotary motion, linear motion, or a combination of both. The most complex devices give motion in three axes and rotations around two of those axes. To generate the mechanical movement inside the chamber, two basic mechanisms are commonly employed: a mechanical coupling through the vacuum wall (using a vacuum-tight seal around the coupling), or a magnetic coupling that transfers motion from air-side to vacuum-side. Various forms of motion control are available for manipulators, such as knobs, handwheels, motors, stepping motors, piezoelectric motors, and pneumatics.

The manipulator or sample holder may include features that allow additional control and testing of a sample, such as the ability to apply heat, cooling, voltage, or a magnetic field. Sample heating can be accomplished by electron bombardment or thermal radiation. For electron bombardment, the sample holder is equipped with a filament which emits electrons when biased at a high negative potential. The impact of the electrons bombarding the sample at high energy causes it to heat. For thermal radiation, a filament is mounted close to the sample and resistively heated to high temperature. The infrared energy from the filament heats the sample.

See also

References

  1. DJ Pacey (2003). W. Boyes, ed. Measurement of vacuum; Chapter 10 in Instrumentation Reference Book (Third ed.). Boston: Butterworth-Heinemann. p. 144. ISBN 0-7506-7123-8.
  2. LM Rozanov & Hablanian, MH (2002). Vacuum technique. London; New York: Taylor & Francis. p. 112. ISBN 0-415-27351-X.
  3. LM Rozanov & Hablanian, MH. Vacuum Technique. p. 95. ISBN 0-415-27351-X.